X-ray Tube Voltage & Current Meter  AB-2015E

image AB-2015E allows the simultaneous measurement of X-ray tube voltage and current for calibration and maintenance of X-ray equipment.

Excellent frequency response covers the full range of X-ray equipment, from inverter to single-phase type.

Features

PC interface / Micro-Processor control / Auto zero-adjustment (One click of button)
Self-checking function / User calibration (Digital calibration)
Digital setting of sampling time (50/60Hz auto switch) and delay time (0~255ms)
Display holding and exposure indicator / Connectable to Model AB-2150D-MⅡ*

* Needs modification and adjustment at manufacturer

Specifications

Measuring range
Tube voltage Rated voltage: Less than 20kV - 150kV
Max. kV on display: 188kV
Tube current-RAD Rated current: 3 - 1500mA
Max. current: More than 1900mA
Tube current-FLU Rated current: 0.03 - 15.00mA
Max. mA on display: More than 19.00mA
Measuring accuracy
Tube voltage Less than 40kV: Less than ±0.7kV
40-100kV: Less than ±1.5%±0.1kV
*Traceable up to 100kV
Tube current Radiography: ±1%±1mA
Fluoroscopy: ±3%±0.01mA
Measuring time
Set range Single phase: 2P, 1P (Auto switch for 50/60 Hz)
Inverter: 1, 2, 4, 8, 16, 32ms
Set accuracy More than 10ms: less than ±2% of set value
Less than 10ms: less than ±200μs
Monitor output
Anode voltage 20kV/V
Cathode voltage 20kV/V  
Tube voltage   20kV/V  
Tube current   Radiography: 200mA/V
Fluoroscopy: 2mA/V  
Data output
Format Serial character data (ASCII Code)  
Data type Tube voltage and tube current (CSV)
Communication type Standard: RS-232C
Protocol: No control sequence
Delimiter: <CR> <LF>  
Communication details Baud rate: 9600 Baud (fixed value)
Data bit length: 8 bits
Parity: N (none)
Stop bit: 1 bit  
General specifications
Power source Voltage: 100V ±10%
Frequency: 50±1 or 60±1 Hz
Power consumption Less than 50 VA (Non-measuring time)
Environment Ambient temperature: 5 - 35℃
Humidity: 45 -85%
Ambient pressure: 70 -106kPa